Real-Time Detection of Circulating Thrombi in an Extracorporeal Circuit Using Doppler Ultrasound: In-Vitro Proof of
Syed Murfad Peer1, Manan Desai2, Syed Bukhari2
1Department of Pediatric Cardiac Surgery, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, USA.
Background:
Thromboembolic stroke continues to be by far the most common severe adverse event in patients supported with mechanical circulatory assist devices. Feasibility of using Doppler ultrasound to detect circulating thrombi in an extracorporeal circuit was investigated.
Methods:
A mock extracorporeal circulatory loop of uncoated cardiopulmonary bypass tubing and a roller pump was setup. A Doppler bubble counter was used to monitor the mean ultrasound backscatter signal (MUBS). The study involved two sets of experiments. In Scenario 1, the circuit was sequentially primed with human blood components, and the MUBS was measured. In Scenario 2, the circuit was primed with heparinized fresh porcine blood, and the MUBS was measured. Fresh blood clots (diameter <1,000 microns, 1,000-5,000 microns, >5,000 microns) were injected into the circuit followed by protamine administration.
Results:
In Scenario 1 (n = 3), human platelets produced a baseline MUBS of 1.5 to 3.5 volts/s. Addition of packed human red blood cells increased the baseline backscatter to 17 to 21 volts/s. Addition of fresh frozen plasma did not change the baseline backscatter. In Scenario 2 (n = 5), the blood-primed circuit produced a steady baseline MUBS. Injection of the clots resulted in abrupt and transient increase (range: 3-30 volts/s) of the baseline MUBS. Protamine administration resulted in a sustained increase of MUBS followed by circuit thrombosis.
Conclusions:
Doppler ultrasound may be used for real-time detection of circulating solid microemboli in the extracorporeal circuit. This technology could potentially be used to design safety systems that can reduce the risk of thromboembolic stroke associated with mechanical circulatory support therapy.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...