Background:
The use of dual mobility (DM) cups has increased quickly. It is hypothesized that femoral neck taper geometry may be involved in the risk of prosthetic impingement and DM cup revision. We aim to (1) explore the reasons for revision of DM cups or head/liners and (2) explore whether certain femoral neck characteristics are associated with a higher risk of revision of DM cups.
Methods:
Primary total hip arthroplasties with a DM cup registered in the Dutch Arthroplasty Register between 2007 and 2021 were identified (n = 7603). Competing risk survival analyses were performed, with acetabular component and head/liner revision as the primary endpoint. Reasons for revision were categorized in cup-/liner-related revisions (dislocation, liner wear, acetabular loosening). Femoral neck characteristics were studied to assess whether there is an association between femoral neck design and the risk of DM cup/liner revision. Multivariable Cox proportional hazard analyses were performed.
Results:
The 5- and 10-year crude cumulative incidence of DM cup or head/liner revision for dislocation, wear, and acetabular loosening was 0.5% (CI 0.4-0.8) and 1.9% (CI 1.3-2.8), respectively. After adjusting for confounders, we found no association between the examined femoral neck characteristics (alloy used, neck geometry, CCD angle, and surface roughness) and the risk for revision for dislocation, wear, and acetabular loosening.
Conclusions:
The risk of DM cup or head/liner revision for dislocation, wear, and acetabular loosening was low. We found no evidence that there is an association between femoral neck design and the risk of cup or head/liner revision.