Related Experiment Video
Updated: Jul 4, 2025

Photoelectron Imaging of Anions Illustrated by 310 Nm Detachment of F−
Published on: July 27, 2018
Solution Cathode Glow Discharge Coupled to Atmospheric Pressure Chemical Ionization for Elemental Detection of S and
Frenio A Redeker1, Kelsey O'Malley1, William P McMahon1
1Department of Chemistry, Georgetown University, 37 and O streets, NW, Washington, DC 20057, USA.
Abstract:
We report a post-plasma chemical ionization approach to evaluate solution cathode glow discharge (SCGD) for S and P elemental analysis. Here, the SCGD serves as a reactor to produce chemical vapors for S and P from organic compounds containing these elements, while a corona discharge operated in negative mode is used to ionize the products. The approach creates long-lived ions in atmospheric pressure, enabling direct investigation of chemical vapor products via mass spectrometric and ion mobility separations. The investigations indicate that SCGD converts S and P to H2SO4 and H3PO4, respectively. These species are then ionized as HSO4HNO3 - and H3PO4NO3HNO3- via reactions with NO3HNO3- produced by corona discharge. The response factors for P among several small molecules varies within 10% of the average response from the compounds, suggesting a reasonable species-independent characteristic. The response factors for S show larger variations among compounds, indicating a higher dependence of chemical vapor generation efficiency on analytes' chemical structures. Detection limits of 15 and 29 ng/mL are achieved for P and S detection, respectively. These figures are limited by background equivalent concentrations and low ion flux in the utilized ion mobility-time of flight mass spectrometer, indicating potential for significant improvements. In particular, the specificity of clustering for S and P-containing ions produced in this approach suggest facile analysis of S and P using quadrupole-based mass spectrometers for improved analytical performance.
More Related Videos
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
08:10Fizzy Extraction of Volatile Organic Compounds Combined with Atmospheric Pressure Chemical Ionization Quadrupole Mass Spectrometry
Published on: July 14, 2017
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Overview