Related Experiment Video
Updated: Jul 4, 2025

Force System with Vertical V-Bends: A 3D In Vitro Assessment of Elastic and Rigid Rectangular Archwires
Published on: July 24, 2018
Pull-out resistance of connected K-wires for osteosynthesis: development of a numerical model
Samuel Berthe1,2, Nadia Bahlouli1, Massamaesso Bilasse1,2
1ICube CNRS UMR7357, Strasbourg University, Strasbourg, France.
Abstract:
A predictive finite element model was developed to investigate the best configuration of a fixation pins system consisting of two K-wires inserted in a synthetic model (Sawbones®) at different angles and secured to a connecting rod. Two key parameters were considered to determine the best configuration delivering the higher pull-out strength and lower pull-out length: the diameter and insertion angle. Results show that as the diameter and insertion angle increased, the pull-out force increased, while the pull-out length decreased. Results are successfully compared with available experimental data in literature. This model can be used as an alternative to experimental study.
Related Concept Videos
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in...
Cable Subjected to a Distributed Load