Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
Published on: April 12, 2011
A new reliable device to assess trunk extensors strength
Waleska Reyes-Ferrada1,2, Luis Chirosa-Ríos1, Ignacio Chirosa-Ríos1
1Department of Physical Education and Sport, Faculty of Sport Sciences, University of Granada, Granada, Spain.
This study found that the functional electromechanical dynamometer (FEMD) is a reliable tool for assessing trunk extensor strength across various velocities and ranges of motion. The device demonstrates stable repeatability and good relative reliability for both concentric and eccentric contractions.
Area of Science:
- Biomechanics
- Kinesiology
- Sports Medicine
Background:
- Trunk extensor strength is crucial for athletic performance and injury prevention.
- Accurate and reliable assessment methods are needed to evaluate this muscle group.
- Functional electromechanical dynamometers (FEMD) offer a potential solution for standardized strength testing.
Purpose of the Study:
- To determine the reliability of trunk extensor strength measurements using a functional electromechanical dynamometer (FEMD).
- To assess the impact of different testing velocities and ranges of motion on measurement reliability.
- To compare the reliability of concentric, eccentric, and isometric contractions.
Main Methods:
- Thirty-one male participants underwent trunk extensor strength assessments.
- Testing was performed at three velocities (V1=0.15, V2=0.30, V3=0.45 m/s) and two ranges of motion (R1=25%, R2=50%), plus isometric contractions at 90°.
- Reliability was quantified using intraclass correlation coefficient (ICC), typical error (TE), and coefficient of variation (CV).
Main Results:
- High absolute reliability (low CV) was observed for average and maximum eccentric strength in the V1R1 condition.
- Relative reliability (good ICC values) was demonstrated for both concentric and eccentric average and maximum strength across tested conditions.
- The most reliable measures for concentric and eccentric strength were average strength in V1R1, and for isometric strength, maximum strength.
Conclusions:
- The functional electromechanical dynamometer (FEMD) is a highly reliable device for assessing trunk extensor strength.
- The findings support the use of FEMD for consistent and dependable evaluation of trunk extensor function in research and clinical settings.
- Specific testing parameters (V1R1) appear optimal for maximizing reliability in certain strength measures.
Related Concept Videos
Design Consideration
The factor of safety is another key aspect...
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as the...
Soundness of Cement
Non-destructive Tests for Concrete Strength
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen also...
Dynamic Modulus of Elasticity of Concrete
The sonic test is a common method to determine the dynamic modulus. In this test, a concrete beam, sized either 6 x 6 x 30 inches or 4 x 4 x 20 inches, is clamped at its center. Vibrations are initiated at one end of the beam by an electromagnetic exciter unit powered by a...