Related Experiment Video
Updated: Jul 4, 2025

A Finite Element Approach for Locating the Center of Resistance of Maxillary Teeth
Published on: April 8, 2020
Assessment of stress shielding around a dental implant for variation of implant stiffness and parafunctional loading
Muhammad Ikman Ishak1, Ruslizam Daud1, Siti Noor Fazliah Mohd Noor2
1Fakulti Kejuruteraan and Teknologi Mekanikal, Universiti Malaysia Perlis (UniMAP), Kampus Alam UniMAP, Pauh Putra, 02600 Arau, Perlis, Malaysia.
Purpose:
The aim of this study was to evaluate the mechanical stimuli transfer at the bone-implant interface via stress and strain energy density transfer parameters. This study also aimed to investigate the effect of different implant stiffness and parafunctional loading values on the defined mechanical stimuli transfer from the implant to the surrounding bone.
Methods:
A three-dimensional finite element model of two-piece threaded dental implant with internal hexagonal connection and mandibular bone block was constructed. Response surface method through face-centred central composite design was applied to examine the influence of two independent factors variables using three levels. The analysis model was fitted to a second-order polynomial equation to determine the response values.
Results:
The results showed that the implant stiffness was more effective than the horizontal load value in increasing the stress and strain energy density transfers. The interaction between both factors was significant in decreasing the likelihood of bone resorption. Decreasing the implant stiffness and horizontal load value led to the increased stress transfer and unexpected decrease in the strain energy density, except at the minimum level of the horizontal load. The increase in the implant stiffness and horizontal load value (up to medium level) have increased the strain energy transfer to the bone.
Conclusions:
The stress and strain energy density were transferred distinctively at the bone-implant interface. The role of both implant stiffness and parafunctional loading is important and should be highlighted in the preoperative treatment planning and design of dental implant.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
Stress Concentrations in Circular Shafts
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Stress Concentrations
The stress...
Applications of Stress
The...
Bearing Stress