Comparison of two different oxygen saturation target ranges for automated oxygen control in preterm infants: a
Fleur Brouwer1, Hylke H Salverda2, Sophie J E Cramer2
1Willem-Alexander Children's Hospital, Department of Paediatrics, Division of Neonatology, Leiden University Medical Center, Leiden, Zuid-Holland, The Netherlands f.brouwer@lumc.nl.
Objective:
To compare the effect of peripheral oxygen saturation (SpO2) target range (TR) (either 91%-95% and 92%-96%) on the frequency and duration of hypoxic and hyperoxic episodes while on automated oxygen control using the OxyGenie controller.
Design:
Randomised cross-over study.
Setting:
Tertiary-level neonatal unit in the Netherlands.
Patients:
Infants (n=27) with a median (IQR) gestational age of 27+0 (25+5-27+3) weeks and postnatal age of 16 (10-22) days, receiving invasive or non-invasive respiratory support.
Interventions:
In both groups supplemental oxygen was titrated to a TR of 91%-95% (TRlow) or 92%-96% (TRhigh) by the OxyGenie controller (SLE6000 ventilator) for 24 hours each, in random sequence. After a switch in TR, a 1-hour washout period was applied to prevent carry-over bias.
Main Outcome Measures:
Frequency and duration of hypoxic (SpO2<80% for ≥1 s) and hyperoxic episodes (SpO2>98% for ≥1 s).
Results:
Hypoxic episodes were less frequent when the higher range was targeted (TRhigh vs TRlow: 2.5 (0.7-6.2)/hour vs 2.4 (0.9-10.2)/hour, p=0.02), but hyperoxic episodes were more frequent (5.3 (1.8-12.3)/hour vs 2.9 (1.0-7.1)/hour, p<0.001). The duration of the out-of-range episodes was not significantly different (hypoxia: 4.7 (2.8-7.1) s vs 4.4 (3.7-6.5) s, p=0.67; hyperoxia: 4.3 (3.3-4.9) s vs 3.9 (2.8-5.5) s, p=0.89).
Conclusion:
Targeting a higher SpO2 TR with the OxyGenie controller reduced hypoxic episodes but increased hyperoxic episodes. This study highlights the feasibility of using an automated oxygen titration device to explore the effects of subtle TR adjustments on clinical outcomes in neonatal care.
Trial Registration Number:
NL9662.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...