External quality assessment survey for SARS-CoV-2 nucleic acid amplification tests in clinical laboratories in Tokyo,
Yoshiyuki Sugishita1, Rie Moriuchi1, Yoshikazu Ishii2
1Quality Control Section, Tokyo Metropolitan Institute of Public Health, 3-24-1 Hyakunin-cho, Shinjuku-ku, Tokyo 169-0073, Japan.
Introduction:
Nucleic acid amplification tests (NAATs) play a pivotal role in clinical laboratories for diagnosing COVID-19. This study aimed to elucidate the accuracy of these tests.
Methods:
In 2021, an external quality assessment of NAATs for SARS-CoV-2 was conducted in 47 laboratories in Tokyo, Japan. In open testing, where the laboratories knew that the samples were intended for the survey, a simulated nasopharyngeal swab suspension sample was used, featuring a positive sample A with a viral concentration of 50 copies/μL, positive sample B with 5 copies/μL, and a negative sample. Laboratories employing real-time RT-PCR were required to report cycle threshold (Ct) values. In blind testing, where the samples were processed as normal test samples, a positive sample C with 50 copies/μL was prepared using a simulated saliva sample.
Results:
Of the 47 laboratories, 41 were engaged in open testing. For sample A, all 41 laboratories yielded positive results, whereas for sample B, 36 laboratories reported positive results, 3 laboratories reported "test decision pending", 1 laboratory reported "suspected positive", and 1 laboratory did not respond. All 41 laboratories correctly identified the negative samples as negative. The mean Ct values were 32.2 for sample A and 35.2 for sample B. In the blind test, six laboratories received samples. Sample C was identified as positive by five laboratories and negative by one laboratory.
Conclusions:
The nature of the specimen, specifically the saliva, may have influenced the blind test outcomes. The identified issues must be meticulously investigated and rectified to ensure accurate results.