Related Experiment Video
Updated: Jul 3, 2025

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Clinical compatibility of magnetic resonance imaging with magnetic intramedullary nails: a feasibility study
Joseph D Femino1, Samuel R Barnes2, Scott C Nelson3
1Department of Orthopaedic Surgery, University of Southern California Keck School of Medicine, 1520 San Pablo St., Suite 2000, Los Angeles, CA, 90033, USA.
Introduction:
The use of magnetic resonance imaging (MRI) with a magnetic intramedullary lengthening nail in place is contraindicated per the manufacturer due to the concern of implant activation and migration. A prior in vitro study did not confirm these complications only noting that a 3.0 T MRI weakened the internal magnet. Therefore, a retrospective analysis of patients who underwent an MRI with a magnetic nail in place was performed to determine if any adverse effects occurred in the clinical setting.
Materials And Methods:
A retrospective review of all patients who underwent an MRI with a magnetic lengthening nail in place was performed. The time spent being imaged in the MRI, number of times the patient entered the MRI suite, and the images obtained were recorded. Radiographs were performed before and after the MRI to determine if any hardware complications occurred. The patients were monitored for any adverse symptoms while they were in the suite.
Results:
A total of 12 patients with 13 nails were identified. Two patients underwent imaging with a 3.0 T MRI while the remaining 10 underwent imaging with a 1.5 T MRI. Each patient entered the MRI suite 2.1 times and spent an average of 84.7 min being imaged in the MRI (range 21-494). No patients noted any adverse symptoms related to the nail while in the suite and no hardware complications were identified.
Conclusion:
MRI appears to be safe with a magnetic nail in place and did not result in any complications. Given the manufacturer's recommendations, informed consent should be obtained prior to an MRI being performed and a 3.0 T MRI should be avoided when possible if further activation of the nail is required.
Insights
Magnetic Resonance Imaging (MRI) is safe with magnetic intramedullary lengthening nails. This study found no adverse effects or hardware complications in patients undergoing MRI with these implants.
Area of Science:
- Orthopedic surgery
- Medical imaging
- Biomedical engineering
Background:
- Manufacturers contraindicate MRI with magnetic intramedullary lengthening nails due to concerns of implant activation and migration.
- A previous in vitro study indicated that 3.0 Tesla (T) MRI weakened the nail's magnet but did not cause activation or migration.
- This study aimed to clinically evaluate the safety of MRI in patients with these nails.
Purpose of the Study:
- To determine the safety and identify any adverse effects of Magnetic Resonance Imaging (MRI) in patients with magnetic intramedullary lengthening nails.
- To clinically validate findings from prior in vitro studies regarding MRI compatibility.
Main Methods:
- A retrospective review was conducted on patients who underwent MRI with a magnetic lengthening nail.
- Data collected included MRI imaging time, number of MRI sessions, and pre/post-MRI radiographs to assess hardware complications.
- Patients were monitored for adverse symptoms during MRI exposure.
Main Results:
- Twelve patients with 13 magnetic nails underwent MRI (10 with 1.5 T, 2 with 3.0 T).
- Patients underwent an average of 2.1 MRI sessions, with an average imaging time of 84.7 minutes.
- No adverse symptoms or hardware complications (implant activation or migration) were observed in any patient.
Conclusions:
- MRI appears to be safe for patients with magnetic intramedullary lengthening nails in place.
- While generally safe, informed consent is recommended, and 3.0 T MRI should be used cautiously if nail activation is intended.
- Clinical findings support the safety of MRI, contradicting initial manufacturer concerns.
More Related Videos
04:42Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:33In Vivo Evaluation of Fracture Callus Development During Bone Healing in Mice Using an MRI-compatible Osteosynthesis Device for the Mouse Femur
Published on: November 14, 2017
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Applications Of NMR In Biology