Related Experiment Video
Updated: Jul 3, 2025

Stereo-Electro-Encephalo-Graphy SEEG With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note
Published on: June 13, 2016
Frame-based versus robot-assisted stereo-electro-encephalography for drug-resistant epilepsy
Chang-Lin Han1, Chien-Chen Chou2,3,4, Hsin-Hung Chen1,3
1Department of Neurosurgery, Neurological Institute, Taipei Veterans General Hospital, Taipei, Taiwan.
Background:
Stereoelectroencephalography (SEEG) is an effective presurgical invasive evaluation for drug-resistant epilepsies. The introduction of robotic devices provides a simplified, accurate, and safe alternative to the conventional SEEG technique. We report our institutional experience with robot-assisted SEEG and compare its in vivo accuracy, operation efficiency, and safety with the more traditional SEEG workflow.
Methods:
All patients with medically refractory focal epilepsy who underwent SEEG depth electrode implantation between 2014 and 2022 were included in this study. Technical advancements of the robot-assisted technique are described. Analyses of patient demographics, electrode implantation accuracy, operation time, and procedure-related complications were performed.
Results:
One hundred and sixty-six patients underwent 167 SEEG procedures. The first 141 procedures were performed using a conventional approach involving a Leksell stereotactic system, and the last 26 procedures were robot-assisted. Among the 1726 depth electrodes that were inserted, the median entry point localization error was as follows: conventional (1.0 mm; range, 0.1-33.5 mm) and robot-assisted (1.1 mm; range, 0-4.8 mm) (P = 0.17). The median target point localization error was as follows: conventional (2.8 mm; range, 0.1-49 mm) and robot-assisted (1.8 mm; range, 0-30.3 mm) (P < 0.001). The median operation time was significantly reduced with the robot-assisted workflow (90 min vs. 77.5 min; P < 0.01). Total complication rates were as follows: conventional (17.7%) and robot-assisted (11.5%) (P = 0.57). Major complication rates were 3.5% and 7.7% (P = 0.77), respectively.
Conclusions:
SEEG is a safe and highly accurate method that provides essential guidance for epilepsy surgery. Implementing SEEG in conjunction with multimodal planning systems and robotic devices can further increase safety margin, surgical efficiency, and accuracy.