Intra-Abdominal Hemorrhage Triggering Inappropriate Therapy From a Subcutaneous Defibrillator
Michael M Malaty1, Max Ray1, David Ferreira1,2
1Department of Cardiology, John Hunter Hospital, New Lambton, New South Wales, Australia.
Abstract:
The SMART Pass filter (Boston Scientific) aims to reduce inappropriate shocks (IASs) from subcutaneous implantable cardioverter-defibrillators by filtering out low-frequency signals such as T waves. However, this filter is deactivated in the presence of diminished R-wave sensing. We describe a case of IAS in the setting of extensive intra-abdominal hemorrhage.