Technical note: Minimizing CIED artifacts on a 0.35 T MRI-Linac using deep learning
Austen N Curcuru1, Deshan Yang2, Hongyu An3
1Department of Radiation Oncology, Washington University in St. Louis, St. Louis, Missouri, USA.
This study used CycleGAN to reduce implantable cardioverter defibrillator (ICD) artifacts in MRI-guided radiotherapy (MRgRT) images. The AI model significantly improved image quality and target tracking accuracy for better patient treatment.
Area of Science:
- Medical Imaging
- Artificial Intelligence
- Radiotherapy
Background:
- Implantable cardioverter defibrillators (ICDs) create artifacts that challenge magnetic resonance imaging (MRI)-guided radiotherapy (MRgRT).
- Artifacts can compromise image quality and treatment accuracy.
Purpose of the Study:
- To evaluate an unsupervised generative adversarial network (CycleGAN) for mitigating ICD artifacts in balanced steady-state free precession (bSSFP) cine MRIs.
- To enhance image quality and tracking performance for MRgRT procedures.
Main Methods:
- A CycleGAN model was trained on bSSFP MRI data from 14 healthy volunteers with simulated ICDs.
- The model was tested using a Leave-One-Out cross-validation scheme.
- Image quality and tracking metrics (DSC, TRE, 95% HD, nRMSE, PSNR, MS-SSIM) were evaluated, alongside qualitative assessment on patient data.
Main Results:
- CycleGAN reconstruction significantly improved whole-heart contour tracking: Dice similarity coefficient (DSC) increased from 0.910 to 0.935, target registration error (TRE) decreased from 4.488 to 2.877 mm, and 95% Hausdorff distance (95% HD) decreased from 10.236 to 7.700 mm.
- Image quality metrics also improved: normalized root mean square error (nRMSE) decreased from 0.644 to 0.420, multiscale structural similarity (MS-SSIM) increased from 0.779 to 0.819, and peak signal-to-noise ratio (PSNR) increased from 18.744 to 22.368.
- Qualitative analysis of additional datasets confirmed a reduction in ICD artifacts.
Conclusions:
- CycleGAN effectively mitigates implantable cardioverter defibrillator (ICD) artifacts in MRI-guided radiotherapy (MRgRT).
- The AI-driven approach significantly enhances both image quality and tracking accuracy, crucial for precise radiotherapy delivery.