Quality control log with CUSUM and clinically useful limits criteria
Archives of Pathology & Laboratory Medicine
|April 1, 1985
Summary
This study introduces a new quality control (QC) log system for clinical chemistry assays. The QC log offers advantages over traditional charts for monitoring analytic drift and improving QC decisions.
Area of Science:
- Clinical Chemistry
- Analytical Science
- Quality Control
Background:
- Traditional quality control charts like Levey-Jennings and Shewart are widely used in clinical chemistry.
- These methods may have limitations in detecting specific types of analytic drift.
- There is a need for practical and effective QC systems in quantitative assays.
Purpose of the Study:
- To describe a novel quality control (QC) log system for quantitative assays in clinical chemistry.
- To highlight the advantages of this QC log over existing graphic charts.
- To provide a practical system for analytic quality control.
Main Methods:
- The QC log system utilizes cumulative sum (CUSUM) techniques to monitor acute analytic drift.
- Comparison of means is employed to detect chronic analytic drift.
- Clinically useful limits (CUL) criteria are integrated for QC decision-making.
- An algorithm is provided to aid in the description, training, and implementation of the system.
Main Results:
- The proposed QC log system offers practical advantages for analytic quality control.
- It effectively monitors both acute and chronic analytic drift.
- The use of CUL criteria enhances the reliability of QC decisions.
- The provided algorithm simplifies the adoption and use of the QC log.
Conclusions:
- The described QC log system is a practical and effective tool for analytic quality control in clinical chemistry.
- It provides superior monitoring capabilities for analytic drift compared to traditional charts.
- The system, incorporating CUSUM, mean comparison, and CUL criteria, improves the accuracy and utility of QC processes.