Quantifying the accuracy of inter-beat intervals acquired from consumer-grade photoplethysmography wristbands using
Xingran Cui1,2, Jing Wang1, Shan Xue1
1School of Biological Science and Medical Engineering, Southeast University, Nanjing, People's Republic of China.
Abstract:
Objective. Although inter-beat intervals (IBI) and the derived heart rate variability (HRV) can be acquired through consumer-grade photoplethysmography (PPG) wristbands and have been applied in a variety of physiological and psychophysiological conditions, their accuracy is still unsatisfactory.Approach.In this study, 30 healthy participants concurrently wore two wristbands (E4 and Honor 5) and a gold-standard electrocardiogram (ECG) device under four conditions: resting, deep breathing with a frequency of 0.17 Hz and 0.1 Hz, and mental stress tasks. To quantitatively validate the accuracy of IBI acquired from PPG wristbands, this study proposed to apply an information-based similarity (IBS) approach to quantify the pattern similarity of the underlying dynamical temporal structures embedded in IBI time series simultaneously recorded using PPG wristbands and the ECG system. The occurrence frequency of basic patterns and their rankings were analyzed to calculate the IBS distance from gold-standard IBI, and to further calculate the signal-to-noise ratio (SNR) of the wristband IBI time series.Main results.The accuracies of both HRV and mental state classification were not satisfactory due to the low SNR in the wristband IBI. However, by rejecting data segments of SNR < 25, the Pearson correlation coefficients between the wristbands' HRV and the gold-standard HRV were increased from 0.542 ± 0.235 to 0.922 ± 0.120 for E4 and from 0.596 ± 0.227 to 0.859 ± 0.145 for Honor 5. The average accuracy of four-class mental state classification increased from 77.3% to 81.9% for E4 and from 79.3% to 83.3% for Honor 5.Significance.Consumer-grade PPG wristbands are acceptable for HR and HRV monitoring when removing low SNR segments. The proposed method can be applied for quantifying the accuracies of IBI and HRV indices acquired via any non-ECG system.
More Related Videos
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
08:22Author Spotlight: Advancing the Study of Brain-Heart Interplay with a Comprehensive EEGLAB Plugin for Multimodal Signal Analysis
Published on: April 26, 2024
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of blood pressure in brachial artery(two-step method)
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....