Calibration of Dual-Channel Raman Spectrometer via Optical Frequency Comb
Shengyujie Lv1, Xiaoping Lou1, Qiaona Gai1
1School of Instrument Science and Opto-Electronics Engineering, Beijing Information Science and Technology University, Beijing 100192, China.
Abstract:
The portable Raman spectrometer boasts portability, rapid analysis, and high flexibility. It stands as a crucial and powerful technical tool for analyzing the chemical composition of samples, whether biological or non-biological, across diverse fields. To improve the resolution of grating spectrometers and ensure a wide spectral range, many spectrometer systems have been designed with double-grating structures. However, the impact of external forces, such as installation deviations and inevitable collisions, may cause differences between the actual state of the internal spectrometer components and their theoretical values. Therefore, spectrometers must be calibrated to establish the relationship between the wavelength and the pixel positions. The characteristic peaks of commonly used calibration substances are primarily distributed in the 200-2000 cm-1 range. The distribution of characteristic peaks in other wavenumber ranges is sparse, especially for spectrometers with double-channel spectral structures and wide spectral ranges. This uneven distribution of spectral peaks generates significant errors in the polynomial fitting results used to calibrate spectrometers. Therefore, to satisfy the calibration requirements of a dual-channel portable Raman spectrometer with a wide spectral range, this study designed a calibration method based on an optical frequency comb, which generates dense and uniform comb-like spectral signals at equal intervals. The method was verified experimentally and compared to the traditional calibration method of using a mercury-argon lamp. The results showed that the error bandwidth of the calibration results of the proposed method was significantly smaller than that of the mercury-argon lamp method, thus demonstrating a substantial improvement in the calibration accuracy.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
UV–Vis Spectrometers
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectroscopy: Beer–Lambert Law
UV–Vis Spectroscopy: Molecular Electronic Transitions