A comparative study on the deformation behavior and mechanical properties of new lower extremity arterial stents
Haiquan Feng1, Xinyuan Shi1, Tianqi Wang1
1College of Mechanical Engineering, Inner Mongolia University of Technology, Hohhot 010000, PR China.
Background And Objective:
The lower extremity movement involves a complex and large amplitude extremity movement process, and arterial stents implanted in the lower extremity are prone to complex mechanical deformation behavior. Hence, the lower extremity arterial stent is required to have favorable comprehensive mechanical properties.
Methods:
In this study, a new lower extremity arterial stent (New) was proposed, and its deformation behavior and mechanical properties were analyzed by numerical simulations under different deformation modes, such as radial compression, axial compression/tension, bending, and torsion. Stents with different diameters were modeled to compare the effect of diameter size on their biomechanical properties. Additionally, a comparative analysis was conducted between this new stent and seven commercially available stents.
Results:
The results demonstrated that the stent diameter exerted a significant effect on its deformation behavior and mechanical properties. Specifically, with the increase of the stent diameter, the radial expansion rate, radial shrinkage rate, radial support stiffness, axial compression stiffness, and axial tensile stiffness tended to decrease, and the expansion inhomogeneity, stenosis rate, bending stiffness, and torsional stiffness tended to increase. In contrast, the stent diameter exerted a small effect on the stent axial shortening rate and ellipticity. The new lower extremity arterial stent was validated to outperform other stents in terms of most performance indicators. Especially, the radial expansion rate and ellipticity of the New stent were better than those of all commercially available stents. Moreover, the New stent presented favorable mechanical properties and flexibility under the premise of ensuring the support performance.
Conclusions:
Based on these findings, this lower extremity arterial stent may play a better therapeutic effect in clinical application. Furthermore, these analysis results may provide reference for the clinical application and selection of the stent.