Related Experiment Video
Updated: Jul 2, 2025

Visualization of the Immunological Synapse by Dual Color Time-gated Stimulated Emission Depletion STED Nanoscopy
Published on: March 24, 2014
Impact of a tilted coverslip on two-photon and STED microscopy
Guillaume Le Bourdellès1, Luc Mercier1, Johannes Roos1
1Univ. Bordeaux, CNRS, IINS, UMR5297, F-33000 Bordeaux, France.
Abstract:
The advent of super-resolution microscopy has opened up new avenues to unveil brain structures with unprecedented spatial resolution in the living state. Yet, its application to live animals remains a genuine challenge. Getting optical access to the brain in vivo requires the use of a 'cranial window', whose mounting greatly influences image quality. Indeed, the coverslip used for the cranial window should lie as orthogonal as possible to the optical axis of the objective, or else significant optical aberrations occur. In this work, we assess the effect of the tilt angle of the coverslip on STED and two-photon microscopy, in particular, image brightness and spatial resolution. We then propose an approach to measure and reduce the tilt using a simple device added to the microscope, which can ensure orthogonality with a precision of 0.07°.
Related Concept Videos
Total Internal Reflection Fluorescence Microscopy
Confocal Fluorescence Microscopy
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Overview of Microscopy Techniques