Pupil size measurements with a multifunctional aberrometer/coherence interferometer/tomographer and two
Petra Davidova1, Marvin Biller, Vanessa Ademmer
1From the Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany .
Purpose:
To compare precision of pupil size measurements of a multifunctional device (Pentacam AXL Wave [Pentacam]) and 2 infrared-based pupillometers (PupilX, Colvard) and to compare repeatability of Pentacam and PupilX.
Setting:
Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
Design:
Prospective, comparative trial.
Methods:
Pupil diameter of healthy eyes was measured with Colvard once and Pentacam without glare (WO) and with glare (WG), PupilX in 0, 1, and 16 lux 3 times each. In a second series, measurements with Pentacam WO and PupilX in 0.06 and 0.12 lux were assessed.
Results:
36 eyes of participants aged 21 to 63 years were included. Mean pupil diameter was 6.05 mm with Colvard, 5.79 mm (first series), 5.50 mm (second series) with Pentacam WO, 3.42 mm WG, 7.26 mm PupilX in 0, 4.67 mm 1, 3.66 mm 16, 6.82 mm in 0.06, and 6.39 mm in 0.12 lux. Measurements with Pentacam WO were significantly different to PupilX in 0, 0.06, 0.12, and 1 lux (all P < .001), but not to Colvard ( P = .086). Pupil size measured with Pentacam WG and PupilX in 16 lux was not significantly different ( P = .647). Consecutive measurements with Pentacam WO and WG had mean SD of 0.23 mm and 0.20 mm, respectively, and with PupilX 0.11 in 0, 0.24 mm 1, and 0.20 mm in 16 lux.
Conclusions:
Pentacam provided good assessment of pupil size but was not equivalent to PupilX in low lighting conditions. Repeatability was more favorable for Pentacam.