Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Subjective and Objective Refraction Using Autorefraction and Two Wavefront-based Aberrometers in
Petra Davidova1, Mark Rostom1, Luca Gabriel Timoceanu1
1From the Department of Ophthalmology, Goethe-University, Frankfurt, Germany (P.D., M.R., L.G.T., T.S., K.P.K., T.J., and T.K.).
Objective refraction (OR) using modern devices does not fully agree with subjective refraction (SR) after extended-depth-of-focus or simultaneous vision intraocular lens implantation. Agreement varied by device and lens type, with pupil size and IOL toricity influencing results.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Engineering
Background:
- Intraocular lenses (IOLs) are crucial for vision correction after cataract surgery.
- Extended-depth-of-focus (EDOF) and simultaneous vision intraocular lenses (SVIOLs) offer advanced visual rehabilitation.
- Accurate refractive outcomes are essential for patient satisfaction and visual function.
Purpose of the Study:
- To compare subjective refraction (SR) with objective refraction (OR) using autorefractors and aberrometers post-EDOF/SVIOL implantation.
- To evaluate the agreement between SR and OR measurements.
- To identify biometric factors influencing refractive outcome discrepancies.
Main Methods:
- Prospective comparative assessment of agreement between SR and OR.
- Objective refraction measured using autorefractor (AR), Hartmann-Shack aberrometer, and ray-tracing aberrometer.
- Analysis of spherical/cylindrical error, spherical equivalent (SE), and vector components.
- Biometric parameters including axial length, pupil size, IOL power, and toricity were assessed.
Main Results:
- Significant differences observed between SR and OR for most refractive parameters, except vector components.
- Intraclass correlation coefficients between SR and OR were generally poor to moderate.
- Ray-tracing aberrometry showed higher agreement (≤0.50 D difference) in the EDOF group (79.5%), while AR was superior in the SVIOL group (96.4%).
- Pupil size and IOL toricity impacted SE differences, particularly with Hartmann-Shack aberrometry.
Conclusions:
- Objective refraction measurements are not interchangeable with subjective refraction after EDOF/SVIOL implantation.
- Agreement between SR and OR was greater for SVIOLs compared to EDOF IOLs.
- Pupil size and IOL toricity are significant influencing factors on refractive outcome discrepancies.