Quantitation of thin-layer chromatograms with an Apple II computer-based videodensitometer
Analytical Biochemistry
|November 1, 1985
Summary
A new, affordable high-speed densitometer was developed using an Apple II computer and readily available components. This system enables rapid thin-layer chromatography (TLC) spot analysis, offering high precision for lipid quantification.
Area of Science:
- Analytical Chemistry
- Chromatography
- Instrumentation
Background:
- Thin-layer chromatography (TLC) is a widely used separation technique.
- Accurate quantification of separated spots is crucial for quantitative TLC.
- Existing densitometry methods can be expensive or time-consuming.
Purpose of the Study:
- To describe the development of an inexpensive, high-speed densitometer for TLC.
- To evaluate the performance and accuracy of the developed densitometer.
- To demonstrate the feasibility of using a personal computer-based system for quantitative analysis.
Main Methods:
- Construction of a densitometer using an Apple II computer, a black-and-white video camera, and an image digitizing board.
- Integration of a fast coprocessor to accelerate image acquisition and processing.
- Calibration using charred lipid standards of known weights (250-2000 ng).
Main Results:
- The system achieves rapid measurement of TLC spots in 30-40 seconds.
- High precision was demonstrated with coefficients of variation around 0.5% for 500-ng lipid spots.
- A strong second-order correlation (R=0.998) was observed between lipid spot weight and density readings.
Conclusions:
- The developed Apple II-based densitometer offers a cost-effective and high-speed solution for quantitative TLC.
- The system requires minimal electronic expertise for assembly from commercial components.
- This approach provides a reliable and accurate method for lipid quantification in TLC analysis.