Precision evaluation of Raman densimetry for carbon dioxide: improvement by correction for the drift effect
Abstract:
To evaluate the precision of the wavenumber difference between the two main Raman bands of C O 2 (designated as Δ, an index of C O 2 density), we performed theoretical calculations for the uncertainties of the peak positions of the C O 2 Raman bands, and measured Raman spectra of C O 2 under the conditions used for the calculations. The Δ precision improves with increased peak intensity, but the peak intensity-precision relation does not hold beyond an intensity threshold corresponding to the measurement time of 1-2 min. Correction for temporal variation of Δ by fitting a linear function to the data distribution improved the Δ precision to ±0.0005c m -1 at best, corresponding to C O 2 density precision as ±0.00015g/c m 3.
More Related Videos
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...