Tomosynthesis of the Appendicular Skeleton on a Twin Robotic X-ray System: A Cadaveric Fracture Study
Theresa Sophie Patzer1, Andreas Steven Kunz1, Henner Huflage1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080 Würzburg, Germany.
Rationale And Objectives:
Aiming to offset image quality limitations in radiographs due to superimposition, this study investigates the diagnostic potential of appendicular skeleton tomosynthesis.
Materials And Methods:
Eight cadaveric extremities (four hands and feet) were examined employing the prototypical tomosynthesis mode of a twin robotic X-ray scanner. 12 protocols with varying sweep angles (10, 20 vs. 40°), frame rates (13 vs. 26 fps), and tube voltages (60 vs. 80 kV) were compared to radiographs. Four radiologists separately evaluated cortical and trabecular bone visualization and fracture patterns. Interreader reliability was assessed based on the intraclass correlation coefficient (ICC).
Results:
Radiation dose in radiography was 0.59 ± 0.20 dGy * cm2 versus 0.11 ± 0.00 to 2.46 ± 0.17 dGy * cm2 for tomosynthesis. Cortical bone display was inferior for radiographs compared to 40° and 20° tomosynthesis. Best results were ascertained for the 80 kV/40°/26 fps protocol. Trabecular bone depiction was also superior in tomosynthesis (p ≤ 0.009) and best with the 80 kV/10°/26 fps setting. Interreader reliability was moderate for cortical bone display (ICC 0.521, 95% confidence interval 0.356-0.641) and good for trabecular bone (0.759, 0.697-0.810). Diagnostic accuracy for articular involvement and multifragment situations was higher in tomosynthesis (93.8-100%/92.2-100%) vs. radiography (85.9%/82.8%.). Diagnostic confidence was also better in tomosynthesis (p ≤ 0.003).
Conclusion:
Compared to radiography, tomosynthesis allows for superior assessability of cortical and trabecular bone and fracture morphology, especially at high framerates. Operating on a multipurpose X-ray system, tomosynthesis of the appendicular skeleton can be performed without additional scanner hardware.
More Related Videos
08:04Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
09:34A Method to Estimate Cadaveric Femur Cortical Strains During Fracture Testing Using Digital Image Correlation
Published on: September 14, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging