Related Experiment Video
Updated: Jul 1, 2025

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Estimating of radiation output of X-RAY tube using mathematical model: Case of high-frequency machines
Alex Eclador Ngankem1, Clemence Raïssa Tchatchouang Simo2, Daniel Bongue1
1Centre for Atomic Molecular Physics and Quantum Optics (CEPAMOQ), University of Douala, P.O 8580 Douala, Cameroun.
Purpose:
During any radiological procedure, it is important to know the dose to be-administered to the patient and this can be done by estimating the output of the X-ray tube either with a dosimeter or with a mathematical equation or Monte Carlo simulations. The aim of this work is to develop a new mathematical model equation (NMME) for estimating the output of high-frequency X-ray tubes.
Methods:
To achieve this, data collected from ten machines in many regions of Cameroon were used (for nine machines) to build an initial model that does not take into account the anode angle and the tenth machine was used to test the model. Using the SpekCalc software, some simulations were carried out to evaluate the influence of the anode angle. This allowed the NMME to be proposed.
Results:
The deviations frequencies between 0.65% and 19.61% were obtained by comparing the output values obtained using initial model with the measured values. The statistical hypothesis test showed that the estimated values using initial model and NMME are in agreement with those measured unlike the Kothan and Tungjai model. For the tenth machine, the percentage difference between estimated and measured values is less than 8 %.
Conclusion:
These results show that the proposed model performed better than the previous models. In the absence of a dosimeter, the NMME could be used to estimate the output of high frequency X-ray machines and therefore the radiation doses received by patients during diagnostic X-ray examinations.
More Related Videos
Related Concept Videos
Simplified Synchronous Machine Model
In this model, each generator is connected to a...
Transformers with Off-Nominal Turns Ratios
Wind Turbine Machine Models
Induction machines interact through the rotating magnetic field generated by the stator and the rotor. The key parameter is slip, which is the difference between synchronous speed and rotor speed relative to synchronous speed. Slip is...
X-ray Imaging
Biological Effects of Radiation
Multimachine Stability
In analyzing the system, the nodal equations represent the relationship between bus voltages, machine voltages, and machine currents. The nodal equation is given by: