External quality assessment for molecular detection of Ureaplasma urealyticum in China
Yanxi Han1, Jian Jiang1, Yu Ma1
1National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/ National Center of Gerontology, P. R. China; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing, P. R. China; Beijing Engineering Research Center of Laboratory Medicine, Beijing Hospital, Beijing, P. R. China.
Background And Aims:
A pilot external quality assessment (EQA) scheme for molecular detection of Ureaplasma urealyticum (UU) was conducted by the National Center for Clinical Laboratories (NCCL) to evaluate the testing capabilities of clinical laboratories and the actual performance of DNA-based nucleic acid amplification tests (NAAT) and RNA-based NAATs when applied in clinical settings.
Materials And Methods:
The EQA panel contained twelve lyophilized samples, including positive samples containing inactivated cell culture supernatants of UU at different concentrations and sterile saline for negative samples. The positive samples were further divided into three groups of high, moderate and low concentrations. The panels were distributed to the participants and the datasets were analyzed according to the qualitative results.
Results:
A total of 365 laboratories participated in the EQA scheme, and 360 results submitted by 338 laboratories were collected, of which 96.11 % (346/360) of the returned results and 95.86 % (324/338) of the laboratories were deemed competent. The positive percentage agreement (PPA) was ≥ 97.5 % for high and moderate concentration samples, but varied significantly for low concentration samples, decreasing from 86.94 % to 51.94 % as the sample concentration decreased. Additionally, for low concentration samples, RNA-based NAAT showed higher PPAs than DNA-based NAATs, but these results were specific to UU supernatants used in this study.
Conclusion:
Most of UU detection assays employed by the participants were generally consistent with their estimated limit of detection (LOD), and the majority of participants can reliably detect UU samples with high and moderate concentrations, while the poor analytical performance for low concentration samples requires further improvement and optimization.