Related Experiment Video
Updated: Jul 1, 2025

08:10
Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
1.7K
Lead-Specific Performance for Atrial Fibrillation Detection in Convolutional Neural Network Models Using Sinus Rhythm
Shinya Suzuki1, Jun Motogi2, Takuya Umemoto2
1Department of Cardiovascular Medicine, The Cardiovascular Institute Tokyo Japan.
Circulation Reports
|March 11, 2024
Summary
A convolutional neural network (CNN) model effectively detects atrial fibrillation (AF) using sinus rhythm ECG (SR-ECG). A double-lead ECG (leads I and V1) demonstrated performance comparable to an 8-lead ECG for AF screening.
Area of Science:
- Cardiology
- Artificial Intelligence in Medicine
- Medical Diagnostics
Background:
- Atrial fibrillation (AF) detection is crucial for preventing stroke.
- Convolutional neural network (CNN) models show promise for AF detection using electrocardiograms (ECGs).
- The optimal ECG lead configuration for CNN-based AF detection using sinus rhythm ECG (SR-ECG) remains to be determined.
Purpose of the Study:
- To evaluate the diagnostic performance of a CNN model for AF detection using various SR-ECG lead configurations.
- To compare the effectiveness of 8-lead, single-lead, and double-lead SR-ECGs for CNN-based AF detection.
- To identify potential alternative ECG lead combinations for efficient AF screening.
Main Methods:
- A retrospective analysis of a prospective cohort study involving 19,170 new patients.
- Development and 5-fold cross-validation of a CNN model using SR-ECGs from 616 AF cases and 3,412 SR cases.
- Evaluation of CNN model performance using 8-lead, single-lead, and double-lead (specifically leads I and V1) SR-ECGs.
Main Results:
- The CNN model achieved an Area Under the Curve (AUC) of 0.872 for the 8-lead ECG.
- A double-lead ECG using leads I and V1 yielded a comparable AUC of 0.871.
- The odds ratio for AF detection was 15.24 for the 8-lead ECG and 14.34 for the double-lead (I, V1) ECG.
Conclusions:
- A CNN model can effectively detect AF using SR-ECG data.
- A double-lead ECG (leads I and V1) provides diagnostic performance comparable to an 8-lead ECG for AF detection.
- The double-lead ECG (I, V1) presents a potential alternative for AF screening, simplifying data acquisition.