Inter-scanner Aβ-PET harmonization using barrel phantom spatial resolution matching
Gihan P Ruwanpathirana1,2, Robert C Williams2, Colin L Masters3,4
1Department of Biomedical Engineering The University of Melbourne Melbourne VIC Australia.
Alzheimer'S & Dementia (Amsterdam, Netherlands)
|March 13, 2024
Summary
Matching spatial resolution between PET scanners using a barrel phantom harmonizes amyloid beta-positron emission tomography (Aβ-PET) measurements. This method ensures consistent Aβ-SUVR quantitation across different scanners, crucial for multi-center trials.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiochemistry
Background:
- Standardized uptake value ratio (SUVR) measures amyloid beta-positron emission tomography (Aβ-PET) uptake in the brain.
- Variability in PET scanner technology and reconstruction techniques causes inconsistencies in Aβ-PET images.
- Harmonization is essential for consistent Aβ-PET measurements across multiple centers and scanners.
Purpose of the Study:
- To propose and validate an Aβ-PET harmonization method based on matching spatial resolution.
- To assess the consistency of Aβ-PET measurements across different PET scanners using phantom-derived spatial resolution.
- To improve the reliability of Aβ-PET quantitation in multi-center clinical trials.
Main Methods:
- Evaluated three PET scanners: Siemens Biograph Vision 600, Siemens Biograph mCT, and Philips Gemini TF64.
- Calculated the full-width at half-maximum (FWHM) using [18F]-filled barrel phantom scans for each scanner's reconstruction configuration.
- Assessed statistical similarity between regional SUVRs using paired t-tests on data from subjects scanned on multiple scanners.
Main Results:
- Specific FWHM values were identified for each scanner pair that maximized statistical similarity in regional SUVRs (e.g., Vision 4.10 mm vs. mCT 4.30 mm).
- Matched barrel phantom spatial resolutions (FWHM) between scanner pairs led to maximal harmonization of Aβ-PET quantitation.
- The method demonstrated effectiveness in harmonizing SUVRs across different scanner types and reconstruction settings.
Conclusions:
- Matching spatial resolution derived from barrel phantom scans effectively harmonizes Aβ-PET quantitation between PET scanner pairs.
- This harmonization technique enhances the consistency and validity of Aβ-PET measurements in multi-center studies.
- The proposed method offers a practical approach for ensuring reliable Aβ-PET data across different imaging sites and over time.