Related Experiment Video
Updated: Jul 1, 2025

Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Adverse events associated with device assisted hyoid and tongue base suspension for obstructive sleep apnea
Jane Y Tong1, Wiktoria A Gocal1, Sunny J Haft1
1Department of Otorhinolaryngology - Head and Neck Surgery, University of Maryland, School of Medicine, Baltimore, MD, United States of America.
Purpose:
Hyoid and tongue base suspension may treat obstructive sleep apnea (OSA). This study summarizes device-related adverse events associated with the AIRvance and AIRLIFT systems used for hyoid and tongue base suspension.
Materials And Methods:
The U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database was queried for reports describing adverse events associated with hyoid or tongue base suspension from January 2012 to December 2022.
Results:
77 adverse events were identified. When performed separately, adverse events were equally as common with hyoid suspension as with tongue base suspension. More complications occurred postoperatively (51 [66.2 %]) than intraoperatively (26 [33.8 %]). The most reported adverse events were infection (23 [29.9 %]), broken screw (15 [19.5 %]), pain or discomfort (10 [13.0 %]), suture rupture (8 [10.4 %]), and dislodged screw (7 [9.1 %]). 10 infections required drainage or debridement; 12 required device explantation.
Conclusions:
The present study is the largest and most longitudinal review of adverse events associated with hyoid and tongue base suspension. Infection was the most common adverse event, and may require device explantation. While adverse events were most frequently attributed to device malfunction, broken screw, suture rupture, and broken needle were often attributed to operator error due to application of excessive force. Surgeon training to increase familiarity with hyoid and tongue base suspension may reduce adverse events caused by operator error. The MAUDE database is limited as a passive surveillance system. Standardized reporting may improve understanding of associated adverse events, enabling better informed comparisons between surgical treatment options for OSA.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Skeletal Muscle Relaxants: Adverse Effects
Unlike...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Suctioning the Nasopharyngeal Airway
Equipment Required