Related Experiment Video
Updated: Jul 1, 2025

Morphological and Functional Assessment of the Right Ventricle Using 3D Echocardiography
Published on: October 28, 2020
Ex vivo evaluation of 3 different right ventricular outflow tract substitutes
Hiroyuki Saisho1, Ioana Geisler1, Michael Scharfschwerdt1
1Department of Cardiac and Thoracic Vascular Surgery, University Hospital of Schleswig Holstein, Lübeck, Germany.
Objectives:
The Ross procedure represents an excellent treatment option in younger patients with aortic stenosis but is limited by poor availability of homografts. In this study, we investigated the hydrodynamic performance of 3 different types of right ventricular outflow tract replacement with pericardium or synthetic material.
Methods:
Three different types of valved conduits were constructed using pericardium and/or synthetic material (Group PEPE: pericardial cusps and pericardial conduit, Group PEPR: pericardial cusps and Dacron conduit, Group PRPR: expanded polytetrafluoroethylene cusps and Dacron conduit). The conduits were designed according to the Ozaki method. Their hydrodynamic performance (effective orifice area, mean pressure gradient and leakage volume) were evaluated in a mock circulation loop at different hydrodynamic conditions.
Results:
Hydrodynamic assessment showed significantly larger effective orifice area of PEPE and PEPR compared to PRPR under all conditions and there were no significant differences between PEPE and PEPR [for condition 2: PEPE 2.43 (2.35-2.54) cm2, PEPR: 2.42 (2.4-2.5) cm2, PRPR: 2.08 (1.97-2.21) cm2, adjusted pairwise comparisons: PEPE versus PEPR: P = 0.80, PEPE versus PRPR: P < 0.001, PEPR versus PRPR: P < 0.001]. Mean pressure gradient was significantly lower for PEPE and PEPR compared with PRPR, whereas no significant differences were seen between PEPE and PEPR. Leakage volume was significantly lower for PEPE and PEPR compared with PRPR under all conditions while leakage was similar between PEPE and PEPR.
Conclusions:
Pulmonary graft reconstruction with pericardium cusps showed superior hydrodynamic performance compared with polytetrafluoroethylene cusps. Our results suggest that it could be considered as an alternative substitute for right ventricular outflow tract replacement during the Ross procedure.
More Related Videos
10:33Assessment of Right Ventricular Structure and Function in Mouse Model of Pulmonary Artery Constriction by Transthoracic Echocardiography
Published on: February 3, 2014
09:40Induction of Right Ventricular Failure by Pulmonary Artery Constriction and Evaluation of Right Ventricular Function in Mice
Published on: May 13, 2019