Related Experiment Video
Updated: Jul 1, 2025

Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
The Absorb GT1 Bioresorbable Vascular Scaffold System - 5-Year Post-Market Surveillance Study in Japan
Masato Nakamura1, Nobuaki Suzuki2, Kenshi Fujii3
1Division of Minimally Invasive Treatment in Cardiovascular Medicine, Toho University Ohashi Medical Center.
Background:
The 1-year clinical outcomes of the Absorb GT1 Japan post-market surveillance (PMS) suggested that an appropriate intracoronary imaging-guided bioresorbable vascular scaffold (BVS) implantation technique may reduce the risk of target lesion failure (TLF) and scaffold thrombosis (ST) associated with the Absorb GT1 BVS. The long-term outcomes through 5 years are now available.
Methods And Results:
This study enrolled 135 consecutive patients (n=139 lesions) with ischemic heart disease in whom percutaneous coronary intervention (PCI) with the Absorb GT1 BVS was attempted. Adequate lesion preparation, imaging-guided appropriate sizing, and high-pressure post-dilatation using a non-compliant balloon were strongly encouraged. All patients had at least 1 Absorb GT1 successfully implanted at the index procedure. Intracoronary imaging was performed in all patients (optical coherence tomography: 127/139 [91.4%] lesions) and adherence to the implantation technique recommendations was excellent: predilatation, 100% (139/139) lesions; post-dilatation, 98.6% (137/139) lesions; mean (±SD) post-dilatation pressure, 18.8±3.5 atm. At 5 years, the follow-up rate was 87.4% (118/135). No definite/probable ST was reported through 5 years. The cumulative TLF rate was 5.1% (6/118), including 2 cardiac deaths, 1 target vessel-attributable myocardial infarction, and 3 ischemia-driven target lesion revascularizations.
Conclusions:
Appropriate intracoronary imaging-guided BVS implantation, including the proactive use of pre- and post-balloon dilatation during implantation may be beneficial, reducing the risk of TLF and ST through 5 years.
More Related Videos
11:12Nondestructive Monitoring of Degradable Scaffold-Based Tissue-Engineered Blood Vessel Development Using Optical Coherence Tomography
Published on: October 3, 2018
09:34Visualizing Angiogenesis by Multiphoton Microscopy In Vivo in Genetically Modified 3D-PLGA/nHAp Scaffold for Calvarial Critical Bone Defect Repair
Published on: September 7, 2017