Determination of meropenem in capillary plasma microsamples using LC-MS/MS
Maria Amélia de Castilhos Busato1,2, Amanda Pacheco Bondan1, Marcos Frank Bastiani1,2
1Laboratório de Toxicologia Analítica, Universidade Feevale, Rua Rubem Berta 200, Novo Hamburgo, Rio Grande do Sul, Brazil.
Bioanalysis
|March 18, 2024
Summary
This study developed a validated LC-MS/MS assay to measure meropenem in capillary plasma microsamples. This method accurately quantifies meropenem, supporting personalized dosing regimens.
Area of Science:
- Pharmacokinetics and Therapeutic Drug Monitoring
- Analytical Chemistry
- Clinical Diagnostics
Background:
- Meropenem plasma concentration measurement is crucial for individualizing dosing regimens.
- Accurate quantification of meropenem is essential for optimizing patient treatment outcomes.
- Existing methods may require larger sample volumes, posing challenges for certain patient populations.
Purpose of the Study:
- To develop and validate a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay.
- To quantify meropenem concentrations in small capillary plasma microsamples.
- To assess the feasibility of using capillary blood for meropenem therapeutic drug monitoring.
Main Methods:
- Sample preparation involved protein precipitation using acetonitrile and dichloromethane clean-up.
- Liquid chromatography-tandem mass spectrometry (LC-MS/MS) was utilized for quantification.
- Assay validation included linearity, accuracy, precision, and matrix effect assessment.
Main Results:
- The developed LC-MS/MS assay demonstrated linearity over the range of 0.5-50 μg/ml.
- Minimal matrix effects were observed, indicating assay robustness.
- High correlation was found between meropenem concentrations in paired capillary and venous plasma samples.
Conclusions:
- A fully validated LC-MS/MS assay for meropenem in capillary plasma microsamples has been established.
- The assay exhibits excellent accuracy and precision, suitable for clinical use.
- This method offers a promising alternative for meropenem therapeutic drug monitoring using minimal sample volumes.