Related Experiment Video
Updated: Jun 30, 2025

Total Internal Reflection Absorption Spectroscopy TIRAS for the Detection of Solvated Electrons at a Plasma-liquid Interface
Published on: January 24, 2018
Crowd control of ions in the Astral analyzer
Hamish Stewart1, Dmitry Grinfeld1, Johannes Petzoldt1
1Thermo Fisher Scientific, Bremen, Germany.
Abstract:
Space charge effects are the Achilles' heel of all high-resolution ion optical devices. In time-of-flight mass analyzers, these may manifest as reduction of resolving power, mass measurement shift, peak coalescence, and/or transmission losses, while highly sensitive modern ion sources and injection devices ensure that such limits are easily exceeded. Space charge effects have been investigated, by experiment and simulation study, for the astral multi-reflection analyzer, incorporating ion focusing via a pair of converging ion mirrors, and fed by a pulsed extraction ion trap. Major factors were identified as the resonant effect between ~103 ions of similar m/z in-flight and the expansion of trapped packets of ~104-5 ions prior to extraction. Optimum operation and compensated ion mirror calibration strategies were then generated and described based on these findings.
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Ion-Exchange Chromatography
Controlled-Current Coulometry: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...