Long-Term Outcomes of the Altis® Single-Incision Sling: Up to 10 Years' Follow-up
Luis Miguel Quintana Franco1, Raquel González López2, Héctor Garde García2
1Department of Urology, Hospital Universitario Fundación Jiménez Díaz, Madrid, Spain. luismi.quintanafranco@gmail.com.
Introduction And Hypothesis:
Single-incision slings (SIS) have emerged as a less invasive alternative to conventional slings for stress urinary incontinence (SUI) treatment. However, long-term efficacy and safety results remain uncertain owing to a lack of studies.
Material And Methods:
A retrospective review of 155 patients treated with Altis® for SUI between February 2012 and June 2017, held in 2022, as a continuation of a prospective study in which all patients (197) were reviewed for 2 years after surgery (1, 6, 12, and 24 months). Preoperative demographic data, comorbidities, and pressure-flow studies were also recorded. Continence status and satisfaction rates were assessed using the International Consultation on Incontinence Questionnaire-short form (ICIQ-SF) and the Patient Global Impression of Improvement (PGI-I) respectively. The assessment in the 2022 retrospective review was performed via a telephone survey.
Results:
Mean follow-up time after surgery was 85.3 months (82.5-88.1). In 2022, complete continence was present in 75.4% of the patients. The presence of urinary urgency conditioned the ICIQ-SF score (10.9 vs 1.7 points, p < 0.01), with the ICIQ-SF = 0 in 84.5% of the patients with no associated urgency. Satisfaction assessed by the PGI-I was high, with 84.6% of the patients showing improvement. De novo urgency was present in 37,9% of the patients by 2022. Urinary tract infections were the most frequent complication (9.7%), with only 5 documented cases of mesh erosion.
Conclusions:
Altis® SIS is a safe and effective device for SUI treatment, with satisfaction rates comparable with those of the conventional slings. Persistence or development of urinary urgency influences the results.