Related Experiment Video
Updated: May 3, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Skin marker combined with surface-guided auto-positioning for breast DIBH radiotherapy daily initial patient setup:
Jianjun Lai1,2, Zhizeng Luo1, Lu Jiang2
1Instiute of Intelligent Control and Robotics, Hangzhou Dianzi University, Hangzhou, China.
Background And Purpose:
By employing three surface-guided radiotherapy (SGRT)-assisted positioning methods, we conducted a prospective study of patients undergoing SGRT-based deep inspiration breath-hold (DIBH) radiotherapy using a Sentine/Catalys system. The aim of this study was to optimize the initial positioning workflow of SGRT-DIBH radiotherapy for breast cancer.
Materials And Methods:
A total of 124 patients were divided into three groups to conduct a prospective comparative study of the setup accuracy and efficiency for the daily initial setup of SGRT-DIBH breast radiotherapy. Group A was subjected to skin marker plus SGRT verification, Group B underwent SGRT optical feedback plus auto-positioning, and Group C was subjected to skin marker plus SGRT auto-positioning. We evaluated setup accuracy and efficiency using cone-beam computed tomography (CBCT) verification data and the total setup time.
Results:
In groups A, B, and C, the mean and standard deviation of the translational setup-error vectors were small, with the highest values of the three directions observed in group A (2.4 ± 1.6, 2.9 ± 1.8, and 2.8 ± 2.1 mm). The rotational vectors in group B (1.8 ± 0.7°, 2.1 ± 0.8°, and 1.8 ± 0.7°) were significantly larger than those in groups A and C, and the Group C setup required the shortest amount of time, at 1.5 ± 0.3 min, while that of Group B took the longest time, at 2.6 ± 0.9 min.
Conclusion:
SGRT one-key calibration was found to be more suitable when followed by skin marker/tattoo and in-room laser positioning, establishing it as an optimal daily initial set-up protocol for breast DIBH radiotherapy. This modality also proved to be suitable for free-breathing breast cancer radiotherapy, and its widespread clinical use is recommended.