Wear and performance of a tripolar total hip replacement
1Seattle, WA, USA.
This study shows that tripolar hip replacements offer low wear and excellent stability, suggesting they are a durable option for high-demand patients. These implants provide a greater range of motion and jump distance compared to conventional and dual mobility prostheses.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- The tripolar hip replacement is an unconstrained prosthesis combining a large-capacity acetabular component with a bipolar prosthesis.
- It aims to prevent dislocation and maintain a near-normal range of motion (ROM), differing from newer dual mobility designs.
- Limited data exists on the long-term wear and performance of tripolar hip replacements.
Purpose of the Study:
- To evaluate the wear, performance, and mechanical function of retrieved tripolar hip prostheses.
- To compare the tripolar prosthesis to conventional and dual mobility designs in high-demand patients.
Main Methods:
- Twenty-four tripolar prostheses were retrieved from 23 patients after a mean of 14 years (range, 5-21 years).
- Implants were from high-demand patients (adventure sports, high-risk occupations, revisions).
- Key design features included a large-diameter, highly cross-linked polyethylene acetabular component, a titanium nitride-coated bipolar, and positive eccentricity.
Main Results:
- Total volumetric wear was 24 mm³/yr, significantly lower than dual mobility (54 mm³/yr) and conventional (38 mm³/yr) prostheses.
- Increased jump distance (16 mm vs. 12 mm) and range of motion (145° vs. 119°) were observed.
- No osteolysis was found, and radiographs indicated continued articulation between components.
Conclusions:
- Tripolar hip prosthesis wear is below the osteolytic threshold, indicating long-term viability.
- Enhanced ROM and jump distance offer superior stability for an unconstrained hip replacement.
- The technology demonstrates safety advantages over dual mobility prostheses, with no intraprosthetic dislocation or adverse metal reactions observed.