Hip resurfacing arthroplasty with cemented highly cross-linked all polyethylene acetabular components
1Department of Orthopedic Surgery, Swedish Medical Center, Seattle, WA, USA.
Annals of Joint
|August 13, 2026
Summary
Cemented highly cross-linked polyethylene (XLPE) acetabular bearings show good survivorship in hip resurfacing arthroplasty (HRA), with a 10-year survival rate of 93.5%. While some wear occurred, it was limited, making XLPE a viable option for HRA.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Arthroplasty Research
Background:
- Hip resurfacing arthroplasty (HRA) is indicated for specific patient groups not well-served by total hip replacement (THR).
- Historically, Metal-on-Metal (MoM) bearings dominated HRA but are now discontinued due to concerns over metal wear debris.
- Highly cross-linked polyethylene (XLPE) bearings have demonstrated excellent outcomes in THR.
Purpose of the Study:
- To evaluate the efficacy and survivorship of cemented highly cross-linked polyethylene (XLPE) acetabular bearings in hip resurfacing arthroplasty (HRA).
- To assess the wear characteristics and clinical outcomes of XLPE bearings in HRA over a long-term follow-up period.
Main Methods:
- A prospective cohort study involving 122 patients undergoing HRA with cemented XLPE acetabular components.
- Patients were followed for a mean of 15 years (range 10-20 years), with clinical outcomes and survivorship meticulously recorded.
- Analysis of retrieved implants (revision or postmortem) was conducted to assess wear debris and implant consequences.
Main Results:
- The 10-year all-cause revision survival rate was 93.5% (95% CI: 91.7-98.1%).
- Six acetabular bearings (4.9%) required revision due to loosening (2), polyethylene wear (2), or dislocation (2).
- Implant retrieval analysis revealed limited anterior/superior wear in most cases, with minimal wear in the dome; two implants showed no signs of wear.
Conclusions:
- Cemented XLPE acetabular bearings demonstrate successful initial clinical outcomes in HRA.
- While effective, cemented XLPE does not offer screw fixation options, limiting its applicability in all HRA cases.