Radiation Exposure and Contrast Agent Use during Endovascular Aortic Repair Using Mobile Versus Fixed Angiography
Amir Arnautovic1, Waseem Garabet1, Reinhold Thomas Ziegler1
1Clinic for Vascular and Endovascular Surgery, Department of Vascular and Endovascular Surgery, Medical Faculty and University Hospital Düsseldorf, Moorenstrasse 5, 40225 Düsseldorf, Germany.
Background:
For (thoracic) endovascular aortic repair ((T)EVAR) procedures, both mobile (standard operating room (SOR)) and fixed C-arm (hybrid operating room (HOR)) systems are available. This study evaluated differences in key procedural parameters, and procedural success for (T)EVAR in the SOR versus the HOR.
Methods:
All patients who underwent standard elective (T)EVAR at the Clinic for Vascular and Endovascular Surgery at the University Hospital Duesseldorf, Germany, between 1 January 2012 and 1 January 2019 were included. Data were retrieved from archived medical records. Endpoints were analyzed for SOR versus HOR during (T)EVAR.
Results:
A total of 93 patients, including 50 EVAR (SOR (n = 20); HOR (n = 30)) and 43 TEVAR (SOR (n = 22); HOR (n= 21)) were included. The dose area product (DAP) for EVAR and TEVAR was lower in the SOR than in the HOR (EVAR, SOR: 1635 ± 1088 cGy·cm2; EVAR, HOR: 7819 ± 8928 cGy·cm2; TEVAR, SOR: 8963 ± 34,458 cGy·cm2; TEVAR, HOR: 14,591 ± 11,584 cGy·cm2 (p < 0.05)). Procedural fluoroscopy time was shorter in the SOR than in the HOR for EVAR and TEVAR (EVAR, SOR: 7 ± 4 min; EVAR, HOR: 18.8 ± 11.3 min; TEVAR, SOR: 6.6 ± 9.6 min; TEVAR, HOR: 13.9 ± 11.8 min (p < 0.05)). Higher volumes of contrast agent were applied during EVAR and TEVAR in the SOR than in the HOR (EVAR, SOR: 57.5 ± 20 mL; EVAR: HOR: 33.3 ± 5 mL (p < 0.05); TEVAR; SOR: 71.5 ± 53.4 mL, TEVAR, HOR: 48.2 ± 27.5 mL (p ≥ 0.05).
Conclusion:
The use of a fixed C-arm angiography system in the HOR results in higher radiation exposure and longer fluoroscopy times but lower contrast agent volumes when compared with mobile C-arm systems in the SOR. Because stochastic radiation sequelae are more likely to be tolerated in an older patient population and, in addition, there is a higher incidence of CKD in this patient population, allocation of patients to the HOR for standard (T)EVAR seems particularly advisable based on our results.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT