Feasibility and Performance of Continuous Glucose Monitoring to Guide Computerized Insulin Infusion Therapy in
Lynn Ang1, Yu Kuei Lin1, Lee F Schroeder2
1Division of Metabolism, Endocrinology & Diabetes, Department of Internal Medicine, University of Michigan, Ann Arbor, MI, USA.
Journal of Diabetes Science and Technology
|April 2, 2024
Summary
Real-time continuous glucose monitoring (CGM) is feasible for managing hyperglycemia in cardiovascular intensive care unit patients. This technology demonstrated accuracy and high satisfaction among nurses and patients, supporting its use in postoperative care.
Area of Science:
- Cardiovascular intensive care
- Endocrinology
- Medical technology
Background:
- Hyperglycemia management in postoperative cardiovascular patients is critical.
- Continuous intravenous insulin infusion (CII) is a common treatment.
- Real-time continuous glucose monitoring (CGM) offers potential for improved titration.
Purpose of the Study:
- To assess the feasibility of CGM for titrating CII in postoperative cardiovascular intensive care unit patients.
- To evaluate the accuracy of CGM compared to point-of-care blood glucose (POC-BG) testing.
- To determine nursing acceptance and patient satisfaction with CGM use.
Main Methods:
- Dexcom G6 CGM devices were used in 59 postsurgical patients with hyperglycemia on CII.
- A hybrid approach combining CGM with periodic POC-BG tests was employed for validation.
- CGM accuracy was evaluated using mean absolute relative difference and Clarke Error Grid analysis.
- Surveys assessed nursing and patient experiences with CGM.
Main Results:
- Analysis of 864 paired POC-BG and CGM values showed mean and median absolute relative differences of 13.2% and 9.8%, respectively.
- 99.7% of paired glucose values fell within acceptable accuracy zones (A and B) on the Clarke Error Grid.
- Nurses found CGM very convenient (93%) and preferred it over POC-BG testing (93%).
- The majority of patients (93%) reported a good or very good care experience with CGM.
Conclusions:
- This pilot study confirms the feasibility, accuracy, and nursing convenience of using CGM with a hybrid approach for insulin titration in postoperative settings.
- The findings support further research into the benefits of CGM for improving workflow, health outcomes, and cost-effectiveness in postoperative care.