Comparison between next-generation sequencing and multiplex polymerase chain reaction assays for nonsmall-cell lung

Shuji Murakami1, Kanako Shinada1, Yuka Otsutsumi2

  • 1Department of Thoracic Oncology, Kanagawa Cancer Center, Yokohama, Kanagawa, Japan.

Cancer Medicine
|April 4, 2024
PubMed
Abstract

Insights

The AmoyDx Pan Lung Cancer PCR Panel (PLC panel) shows a high success rate for detecting mutations in non-small-cell lung cancer (NSCLC). Its performance is comparable to the Oncomine Dx Target Test (ODxTT), offering a valuable diagnostic tool for NSCLC treatment.

Area of Science:

  • Oncology
  • Genetics
  • Molecular Diagnostics

Background:

  • Genetic mutation detection is crucial for targeted therapy in non-small-cell lung cancer (NSCLC).
  • Multiplex tests are needed to identify multiple gene mutations simultaneously for effective treatment planning.
  • Current methods include next-generation sequencing (NGS) and polymerase chain reaction (PCR)-based tests.

Purpose of the Study:

  • To evaluate and compare the performance of an NGS-based test, the Oncomine Dx Target Test (ODxTT), and a real-time PCR-based multiplex test, the AmoyDx Pan Lung Cancer PCR Panel (AmoyDx PLC panel).
  • To assess the success and detection rates of both multiplex biomarker panel tests in NSCLC patients.

Main Methods:

  • A study was conducted on patients with histologically diagnosed NSCLC.
  • Both the AmoyDx PLC panel and ODxTT-M were performed simultaneously on patient samples.
  • Success rates and detection rates for driver mutations, including EGFR mutations, were evaluated.

Main Results:

  • The AmoyDx PLC panel achieved a 100% success rate, while ODxTT-M had a 98% success rate.
  • Nine driver mutations were detected in 35.9% of samples by the AmoyDx PLC panel and 37.3% by ODxTT-M.
  • Concordant results were observed in most cases, with discrepancies mainly attributed to differences in test sensitivity and targeted gene variants.

Conclusions:

  • The AmoyDx PLC panel demonstrates a high success rate and comparable mutation detection frequency to ODxTT-M for key genes in NSCLC.
  • Clinicians should be aware of the specific performance characteristics and limitations of both the AmoyDx PLC panel and ODxTT-M for informed clinical decision-making.
  • Both PCR-based and NGS-based multiplex tests are valuable tools for genetic mutation detection in NSCLC treatment.