Validation of a Suprasystolic Cuff System for Static and Dynamic Representation of the Central Pressure Waveform
Alessio Tamborini1, Morteza Gharib1
1Department of Medical Engineering California Institute of Technology Pasadena CA USA.
Background:
Noninvasive pulse waveform analysis is valuable for central cardiovascular assessment, yet controversies persist over its validity in peripheral measurements. Our objective was to compare waveform features from a cuff system with suprasystolic blood pressure hold with an invasive aortic measurement.
Methods And Results:
This study analyzed data from 88 subjects undergoing concurrent aortic catheterization and brachial pulse waveform acquisition using a suprasystolic blood pressure cuff system. Oscillometric blood pressure (BP) was compared with invasive aortic systolic BP and diastolic BP. Association between cuff and catheter waveform features was performed on a set of 15 parameters inclusive of magnitudes, time intervals, pressure-time integrals, and slopes of the pulsations. The evaluation covered both static (subject-averaged values) and dynamic (breathing-induced fluctuations) behaviors. Peripheral BP values from the cuff device were higher than catheter values (systolic BP-residual, 6.5 mm Hg; diastolic BP-residual, 12.4 mm Hg). Physiological correction for pressure amplification in the arterial system improved systolic BP prediction (r2=0.83). Dynamic calibration generated noninvasive BP fluctuations that reflect those invasively measured (systolic BP Pearson R=0.73, P<0.001; diastolic BP Pearson R=0.53, P<0.001). Static and dynamic analyses revealed a set of parameters with strong associations between catheter and cuff (Pearson R>0.5, P<0.001), encompassing magnitudes, timings, and pressure-time integrals but not slope-based parameters.
Conclusions:
This study demonstrated that the device and methods for peripheral waveform measurements presented here can be used for noninvasive estimation of central BP and a subset of aortic waveform features. These results serve as a benchmark for central cardiovascular assessment using suprasystolic BP cuff-based devices and contribute to preserving system dynamics in noninvasive measurements.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure