Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control
Lukas O Langanky1, Karen B Kreutzer1, Christian F Poets1
1Department of Neonatology, University Children's Hospital, Tübingen University Hospital, Tübingen, Germany.
Abstract:
The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO2) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis of a randomized crossover study of AOC, programmed to set FiO2 to "back-up FiO2" during SL. In 24 preterm infants (median (interquartile range)) gestational age 25.3 (24.6 to 25.6) weeks, recording time 12.7 h (12.2 to 13.6 h) per infant, we identified 76 pHEs (median duration 119 s (86 to 180 s)). In 50 (66%) pHEs, SL occurred for a median duration of 51 s (33 to 85 s) and at a median frequency of 2 (1 to 2) SL-periods per pHE. SpO2 before and after SL was similar (82% (76 to 88%) vs 82% (76 to 87%), p = 0.3)). Conclusion: SL is common during pHE and must hence be considered in AOC-algorithm designs. Administering a "backup FiO2" (which reflects FiO2-requirements during normoxemia) during SL may prolong pHE with SL. Trial registration: The study was registered at www.
Clinicaltrials:
gov under the registration no. NCT03785899.
What Is Known:
• Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range.
What Is New:
• Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...