Universal 120-kV Dual-Source Ultra-High Pitch Protocol on the Photon-Counting CT System for Pediatric Abdomen of All
Wei Zhou1, Donglai Huo, Lorna P Browne
1From the Department of Radiology, University of Colorado, Anschutz Medical Campus, Aurora, CO (W.Z., D.H., L.P.B., J.W.); Department of Radiology, Children's Hospital Colorado, Aurora, CO (L.P.B., J.W.); and Bioinformatics and Computational Biology, University of Minnesota, St Paul, MN (X.Z.).
Objectives:
The purpose of this study is to determine if a universal 120-kV ultra-high pitch and virtual monoenergetic images (VMIs) protocol on the photon-counting computed tomography (PCCT) system can provide sufficient image quality for pediatric abdominal imaging, regardless of size, compared with protocols using a size-dependent kV and dual-source flash mode on the energy-integrating CT (EICT) system.
Materials And Methods:
One solid water insert and 3 iodine (2, 5, 10 mg I/mL) inserts were attached or inserted into phantoms of variable sizes, simulating the abdomens of a newborn, 5-year-old, 10-year-old, and adult-sized pediatric patients. Each phantom setting was scanned on an EICT using clinical size-specific kV dual-source protocols with a pitch of 3.0. The scans were performed with fixed scanning parameters, and the CTDI vol values of full dose were 0.30, 0.71, 1.05, and 7.40 mGy for newborn to adult size, respectively. In addition, half dose scans were acquired on EICT. Each phantom was then scanned on a PCCT (Siemens Alpha) using a universal 120-kV protocol with the same full dose and half dose as determined above on the EICT scanner. All other parameters matched to EICT settings. Virtual monoenergetic images were generated from PCCT scans between 40 and 80 keV with a 5-keV interval. Image quality metrics were compared between PCCT VMIs and EICT, including image noise (measured as standard deviation of solid water), contrast-to-noise ratio (CNR) (measured at iodine inserts with solid water as background), and noise power spectrum (measured in uniform phantom regions).
Results:
Noise at a PCCT VMI of 70 keV (7.0 ± 0.6 HU for newborn, 14.7 ± 1.6 HU for adult) is comparable ( P > 0.05, t test) or significantly lower ( P < 0.05, t test) compared with EICT (7.8 ± 0.8 HU for newborn, 15.3 ± 1.5 HU for adult). Iodine CNR from PCCT VMI at 50 keV (50.8 ± 8.4 for newborn, 27.3 ± 2.8 for adult) is comparable ( P > 0.05, t test) or significantly higher ( P < 0.05, t test) to the corresponding EICT measurements (57.5 ± 6.7 for newborn, 13.8 ± 1.7 for adult). The noise power spectrum curve shape of PCCT VMI is similar to EICT, despite PCCT VMI exhibiting higher noise at low keV levels.
Conclusions:
The universal PCCT 120 kV with ultra-high pitch and postprocessed VMIs demonstrated equivalent or improved performance in noise (70 keV) and iodine CNR (50 keV) for pediatric abdominal CT, compared with size-specific kV images on the EICT.