Settling effect of abutment and torque loss in different types of abutment after cyclic loading
Qinyuan Shen1, Min Huang1, Junchao Wei1
1The Affiliated Stomatological Hospital of Nanchang University & The Key Laboratory of Oral Biomedicine, Jiangxi Province & Jiangxi Province Clinical Research Center for Oral Diseases, Nanchang 330006, China.
Objectives:
This study aimed to evaluate and compare the settling effect of implant-abutment assembly and the torque loss before and after cyclic loading in three types of abutments.
Methods:
Thirty internal hexagon fixtures were randomly divided into three groups (n=10). Group A used original abutments, group B used pre-machined cast abutments, and group C used compatible abutments. In addition, the abutment morse taper angle was measured using an image measuring instrument. Removal torque values (RTVs) were recorded using a digital torque meter before and after cyclic loading. All samples were tested in a universal testing machine with a vertical load between 0 and 250 N for 100 000 cycles of 10 Hz. The settling effect was measured after cyclic loading. Paired t test was performed for intragroup analysis of removal torque loss before and after cycling and one-way ANOVA. Subsequently, Tukey's honestly significant difference test was used for intergroup comparison (α=0.05).
Results:
The paired t-test showed signi⁃ficant differences in the intragroup RTVs before and after cycling (P<0.001). ANOVA showed significant differences in the mean of removal torque loss after cycling (P=0.009), the abutment morse taper angle (P<0.001), and the settling values (P=0.01) among different groups. However, no significant difference was found between compatible abutments and pre-machined cast abutments.
Conclusions:
The screw removal torque was significantly reduced for all groups in this study after cyclic loading. Differences could be found in the internal accuracy of implant-abutment assembly among different groups. Within the limitations of this study, the results showed the original abutments exhibited lower percentages of torque reduction after cyclic loading than the casting abutments and the compatible abutments.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Related Concept Videos
Residual Stresses in Circular Shafts
Eccentric Axial Loading in a Plane of Symmetry
Stress Concentrations in Circular Shafts
Circular Shafts - Elastoplastic Materials
As torque on the...
Thin-Walled Hollow Shafts
Screw: Problem Solving
To evaluate the...