Time difference detection based on sliding window all-phase FFT and Kalman filtering for precise flow measurement
Jiaqi Jing1, Dezhi Zheng2,3, Shangchun Fan1
1School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing 100091, People's Republic of China.
Abstract:
Coriolis mass flowmeter (CMF) measures the mass flow rate by detecting the time difference, typically using frequency domain methods. However, the spectrum leakage is the primary challenge. To address this issue, a new time difference detection method is proposed utilizing sliding window and all-phase fast Fourier transform. The computational complexity is reduced considering the changes in signal frequency. To further improve the stability and response speed of the measured value, a Kalman filtering algorithm based on variance detection is also proposed for post-processing. A transmitter system is developed to validate the proposed methods. The results demonstrate that, for single phase fluids, the accuracy is better than 0.5‰ and the repeatability is better than 0.2‰, thereby offering an improvement in the accuracy of CMF and supporting for industrial applications.