Is the use of gonad protection protectors necessary during infants chest radiography?
1Departamento de Tecnología Radiológica, Facultad de Ciencia y Tecnología de la Salud, Universidad de Bienestar Médico de Kawasaki, Kurashiki, Okayama, Japan.
Introduction And Objectives:
To compare gonad doses with and without a gonad protector and to optimize the use of gonadal protectors in infants thorax radiography.
Materials And Methods:
Two pediatric anthropomorphic phantoms are used: an X-ray system for KXO-50SS/DRX-3724HD, and a digital radiography system for CALNEO Smart C12, with and without a gonad protector during infants thorax radiography. A real time skin dosimeter is placed on the X-ray system, and a real time skin dosimeter is inserted on the front side of the mammary gland, the front and back sides of the true pelvis level, and on the ovaries and testes. The X-ray system is irradiated 15 times using phantoms with and without a gonad protector. The measured entrance patient doses values of for the real time skin dosimeter are compared for each phantom, with and without the gonad protector.
Results:
The medium of measured entrance patient doses values for front side dose of the true pelvis level with and without the protector are 10.00 and 5.00 μGy at newborn, and 10.00 and 0.00μGy at one year, respectively. The medium of measured entrance patient doses values for the back side dose of the true pelvis level with and without the protector are 0.00 and 0.00 μGy at both newborn one year, respectively. The measured entrance patient doses cannot be detected in the ovaries and testes with or without the protector. No significant differences are observed in the measured entrance patient doses values for the front and back side doses of the pelvis, ovaries, and testes at newborn and one year, with and without the protector (p>0.05).
Conclusions:
No significant difference was observed in gonad dose measurements with and without the gonad protector during infants chest radiography. We believe that gonadal protector wearing is not necessary.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging