Related Experiment Video
Updated: Jun 28, 2025

Ultrasensitive Detection of Biomarkers by Using a Molecular Imprinting Based Capacitive Biosensor
Published on: February 16, 2018
Electric Field-Assisted Molecularly Imprinted Polymer-Modified QCM Sensor for Enhanced Detection of Immunoglobulin
Iliya LariMojarad1, MirBehrad Mousavi1, Mohammad Mahdi Moeini Manesh1
1MEMS Lab, School of Electrical and Computer Engineering, College of Engineering, University of Tehran, Tehran 1439957131, Iran.
Abstract:
In this study, an electric-field-assisted molecularly imprinted polymer (EFAMIP) as an enhanced form of MIP was developed to improve the MIP-modified quartz crystal microbalance (QCM) biosensors. While exerting a vertical electric field, polymerization of methacrylic acid in the presence of immunoglobulin G (IgG) as the template was initiated, and later, after the template removal process, the EFAMIPs were obtained. The polymer surface characterization was conducted by using a scanning electron microscope. The impact of electric field direction on IgG binding sites, forming either EFAMIP-Fab or EFAMIP-Fc, was assessed. Next, the static measurement results in liquid for EFAMIP-modified QCM and MIP-modified QCM were compared. While encompassing IgG, EFAMIP-modified QCMs exhibited up to a 113.5% higher frequency shift than typical MIP in time-limited detection. The final frequency shift of EFAMIP, which determines the detection limit of IgG, was improved up to 12.5% compared to typical MIP. Moreover, the EFAMIP-Fab performance was promising for the selective detection of IgG in a solution containing different types of immunoglobulins.