Related Experiment Video
Updated: Jun 28, 2025

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
A novel negative pressure isolation device reduces aerosol exposure: A randomized controlled trial
John S Shin1, Muchuan Niu2, Haoxuan Chen2
1Department of Anesthesiology and Perioperative Medicine, University of California Los Angeles (UCLA), Los Angeles, CA, USA.
Background And Aim:
The COVID-19 pandemic has led to a proliferation of intubation barriers designed to protect healthcare workers from infection. We developed the Suction-Assisted Local Aerosol Containment Chamber (SLACC) and tested it in the operating room. The primary objectives were to determine the ease and safety of airway management with SLACC, and to measure its efficacy of aerosol containment to determine if it significantly reduces exposure to health care workers.
Methods:
In this randomized clinical trial, adult patients scheduled to undergo elective surgery with general endotracheal anesthesia were screened and informed consent obtained from those willing to participate. Patients were randomized to airway management either with or without the SLACC device. Patients inhaled nebulized saline before and during anesthesia induction to simulate the size and concentration of particles seen with severe symptomatic SARS-CoV-2 infection.
Results:
79 patients were enrolled and randomized. Particle number concentration (PNC) at the patients' and healthcare workers' locations were measured and compared between the SLACC vs. control groups during airway management. Ease and success of tracheal intubation were recorded for each patient. All intubations were successful and time to intubation was similar between the two groups. Healthcare workers were exposed to significantly lower particle number concentrations (#/cm3) during airway management when SLACC was utilized vs. control. The particle count outside SLACC was reduced by 97% compared to that inside the device.
Conclusions:
The SLACC device does not interfere with airway management and significantly reduces healthcare worker exposure to aerosolized particles during airway management.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required