Use of a Linear Accelerator for Conducting In Vitro Radiobiology Experiments
Published on: May 26, 2019
Assessment of radio-activation using spectroscopy in medical linear accelerators
Na Hye Kwon1, Young Jae Jang2,3, Suah Yu3,4
1Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic of Korea.
Abstract:
In radiotherapy, when photon energy exceeding 8 MV is utilized, photoneutrons can activate the components within the gantry of the linear accelerator (linac). At the end of the linac's lifecycle, radiation workers are tasked with its dismantling and disposal, potentially exposing them to unintentional radiation. This study aims to identify and measure the radioisotopes generated by this activation through spectroscopy, and to evaluate the effective dose rate. We selected nine medical linacs, considering various factors such as manufacturer (Siemens, Varian, and Elekta), model, energy, period of operation, and workload. We identified the radionuclides in the linac head by employing an in situ high-purity germanium (HPGe) detector. Spectroscopy and dose-rate measurements were conducted post-shutdown. We also measured the dose rates at the beam-exit window following irradiation with 10 MV and 15 MV photon beams. As a result of the spectroscopy, we identified approximately 20 nuclides including those with half-lives of 100 days or longer, such as 54Mn, 60Co, 65Zn, 122Sb, and 198Au. The dose rate measurements after 10 MV irradiation decreased to the background level in 10 min. By contrast, on 15 MV irradiation, the dose rate was 628 nSv/h after 10 min and decreased to 268 nSv/h after 1.5 hours. It was confirmed that the difference in the level of radiation and the type of nuclide depends on the period of use, energy, and workload. However, the type of nuclide does not differ significantly between the linacs. It is necessary to propose appropriate guidelines for the safety of workers, and disposal/move-install should be planned while taking into consideration the equipment's energy usage rate.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Overview
Biological Effects of Radiation
Atomic Fluorescence Spectroscopy
Atomic Emission Spectroscopy: Lab
Atomic Spectroscopy: Absorption, Emission, and Fluorescence