Shielding Performance and Clinical Applicability of Lead-Free Materials in Computed Tomography
Jung Ju Jo1, Kum-Bae Kim1, Vannyat Ath1
1University of Science and Technology, University of Science and Technology (UST), 217 Gajeong-ro, Yuseong-gu, Daejeon 34113, Republic of Korea, Daejeon, 34113, Korea (the Republic of).
None:
Owing to the high radiation exposure associated with computed tomography (CT) examinations and the image quality degradation caused by conventional radiation shielding materials, this study evaluated the dose reduction performance and image quality maintenance potential of a newly developed lead-free composite shielding material. This material was composed of bismuth, tungsten, tungsten carbide, aluminum, and polyurethane. Phantom-based dose measurements demonstrated that the shielding material achieved dose reduction rates ranging from 17.6% to 37.6%, depending on tube voltage. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and changes in tube current-time product (mAs) under a scout-based automatic exposure control (AEC) protocol were analyzed according to the presence or absence of the shielding material across regions. For the clinical evaluation, CT scans were performed on four patients. Furthermore, the images were reviewed to evaluate whether this material affected image quality. The shielding material exhibited radiation reduction levels comparable to those reported in previous studies. SNR and CNR analyses showed minor statistical variations in certain regions; however, most differences were not statistically significant, and even significant differences remained within a range that did not compromise diagnostic image quality. Under the scout-based AEC protocol, the use of the shielding material resulted in less than 1% variation in mAs values. No visually perceptible artifacts or clinically significant image quality degradation were observed. The proposed composite shielding material demonstrated the potential to mitigate some limitations of conventional shielding materials and showed preliminary clinical feasibility as an adjunctive strategy for radiation dose reduction in CT examinations.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
11:31Using Micro-computed Tomography for the Assessment of Tumor Development and Follow-up of Response to Treatment in a Mouse Model of Lung Cancer
Published on: May 20, 2016
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography