Related Experiment Video
Updated: Jun 28, 2025

Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Surgical explantation for a rapid deployment valve: the 'shave-and-hook technique'
Kyokun Uehara1, Taku Shirakami1, Yoshio Arai1
1Department of Cardiovascular Surgery, Tenri Hospital, Nara, Japan.
Abstract:
An 80-year-old female patient underwent redo aortic valve replacement for haemolysis caused by moderate paravalvular leakage 1 year after a 21-mm Intuity Elite valve implantation. The elevatorium passed at the segment with paravalvular leakage. The frame was then bent inward using a hook and the peel around the sawing ring was shaved by an elevatorium. After explantation of the Intuity Elite valve, endoscopic examination showed no sign of annular or sub-annular damage. Conventional aortic valve replacement using a biological valve was performed. We introduce a safe alternative technique for explantation of a rapid deployed valve.