Comparison of Intraocular Lens Power Prediction Accuracy Between 2 Swept-Source Optical Coherence Tomography Biometry
Nir Sorkin1, Rotem Zadok2, Giacomo Savini3
1Department of Ophthalmology, Tel Aviv Medical Center, Tel Aviv, Israel; Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
The Eyestar 900 and IOLMaster 700 offer similar intraocular lens (IOL) power prediction accuracy using the Barrett Universal II formula. Measured posterior keratometry may require lens factor adjustment for both devices.
Area of Science:
- Ophthalmology
- Biomedical Engineering
Background:
- Accurate intraocular lens (IOL) power prediction is crucial for successful cataract surgery outcomes.
- Posterior corneal power significantly influences IOL power calculations.
Purpose of the Study:
- To compare the intraocular lens (IOL) power prediction accuracy of the Eyestar 900 (EyeS900) and IOLMaster 700 (IOLM700).
- To evaluate the impact of estimated posterior keratometry (E-PK) versus measured posterior keratometry (M-PK) on IOL power prediction using the Barrett Universal II (BUII) formula.
Main Methods:
- A retrospective study involving 225 eyes of 225 cataract surgery patients.
- Preoperative measurements were obtained using both the EyeS900 and IOLM700 devices.
- Spherical Equivalent Prediction Error (SEQ-PE), precision, and accuracy were analyzed using the BUII formula with E-PK and M-PK.
Main Results:
- Both EyeS900 and IOLM700 demonstrated comparable IOL power prediction accuracy and precision.
- Statistically significant differences in SEQ-PE were observed between E-PK and M-PK, with M-PK calculations resulting in hyperopic shifts.
- Excellent correlation and agreement were found between the two devices for both E-PK and M-PK.
Conclusions:
- The Eyestar 900 and IOLMaster 700 exhibit similar IOL power prediction accuracy with the BUII formula, regardless of using E-PK or M-PK.
- An adjusted lens factor may be necessary for the BUII formula when employing M-PK with either device.